5 Methods
5.4 General settings
80
Biacore T100 Software Handbook BR-1006-48 Edition AE
5.4
General settings
The settings that are specified here are:
Data collection rate
Choose between 1 and 10 Hz for data collection. The higher setting will provide
better resolution for kinetic analysis of fast interaction processes, but will result
in larger result files.
Detection
Choose the detection setting for the run:
This setting affects the choice of flow path that can be made for each injection
command in the cycle types definitions (see Section 5.6.1). Use the setting
Multi
if you are not sure what you need: this will provide maximum flexibility for the
injection settings.
Sample compartment temperature
This is the temperature in the sample compartment (not the analysis
temperature at the flow cell, which is set for each assay step). Check the
Vary
with analysis temperature
box to set the sample compartment temperature
automatically to the same value as the analysis temperature.
Single
Records data from one flow cell according to the chosen flow path.
Data is not recorded from the other flow cells.
Dual
Records data from one flow cell pair (1,2 or 3,4) according to the
chosen flow path. Data is not recorded from the other flow cell pair.
Multi
Records data from all four flow cells.
Summary of Contents for Biacore T100
Page 1: ...GE Healthcare Biacore T100 Software Handbook ...
Page 2: ......
Page 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Page 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Page 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Page 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 237: ......