8 Concentration analysis
8.2 Evaluating calibrated concentration analyses
144
Biacore T100 Software Handbook BR-1006-48 Edition AE
8.2.4
Custom models for calibration curves
You can define your own fitting models for calibration curves in concentration
analysis, for example to support validated assay procedures that do not use
linear or 4-parameter calibration fitting. Choose
Tools:Models:Concentration
to define or edit fitting models. Models for calibration curves are defined using
similar principles as equation models for kinetics and affinity (see Section 9.9.3).
8.2.5
Evaluating combined result sets
If you use the
Append File
function to combine result sets from separate runs,
concentration analysis can be evaluated provided that the conditions specified
in Section 5.10 are fulfilled in the combined set. The software does not check the
validity of any evaluation applied to a combined result set, so that it is your
responsibility to determine that the evaluation results are meaningful. It is for
example in principle possible to append a kinetic analysis result file to a
concentration analysis, and then apply concentration analysis evaluation:
Summary of Contents for Biacore T100
Page 1: ...GE Healthcare Biacore T100 Software Handbook ...
Page 2: ......
Page 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Page 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Page 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Page 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 237: ......