Kinetics and affinity analysis 9
f
Flow rate (µl/min)
Provided as
input
tOn
Sample injection start time (s)
Provided as
input
tOff
Sample injection end time (s)
Provided as
input
RI
Bulk refractive index contribution in the
sample
Fitted
Report parameters
Calculated as
ka1
Association rate constant for analyte binding
(M
-1
s
-1
)
ka1
kd1
Dissociation rate constant for analyte from
the complex (s
-1
)
kd1
ka2
Forward rate constant for the conformational
change (s
-1
)
ka2
kd2
Reverse rate constant for the conformational
change (s
-1
)
kd2
KD
Overall equilibrium dissociation constant (M)
kd1/ka1
×
(kd2/(kd2+ka2))
Rmax
Analyte binding capacity of the surface (RU)
Rmax
Conc
Analyte concentration (M)
Conc
tc
Flow rate-independent component of the
mass transfer constant
tc
Flow
Flow rate (µl/min)
f
kt
Mass transfer constant
tc
×
f1/3
RI
Bulk refractive index contribution in the
sample
RI
Summary of Contents for Biacore T100
Page 1: ...GE Healthcare Biacore T100 Software Handbook ...
Page 2: ......
Page 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Page 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Page 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Page 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 237: ......