Appendix B Method examples and recommendations
B.2 Calibration-free concentration analysis
218
Biacore T100 Software Handbook BR-1006-48 Edition AE
Blank cycles
For assay step
Blank
, only the flow rate is set at run time. The method variable
Blank
is set to
Yes
in the method and
Sample solution
to
Buffer
. The remaining
variables are not used and may be left blank.
B.2.4
Setup Run
At run time, variables need to be assigned for assay steps
Sample
and
Blank
.
Each sample should be run at two or more flow rates: values of 5 and 100 µl/min
are recommended and are entered for one sample. Use the same set of flow
rates for each sample. Values are also required for analyte molecular weight
(
MW
), diffusion coefficient at 20°C (
D(20°C
) and dilution factor (
Dilution
).
Only flow rates and dilution factor are required for assay step
Blank
: these
values must be the same as the flow rates used for samples.
Summary of Contents for Biacore T100
Page 1: ...GE Healthcare Biacore T100 Software Handbook ...
Page 2: ......
Page 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Page 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Page 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Page 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 237: ......