Biacore T100 Software Handbook BR-1006-48 Edition AE
223
Method examples and recommendations Appendix B
B.5
Kinetics heterogeneous analyte
This method is a straightforward kinetics determination with evaluation
variables included for the evaluation purpose
Kinetics – heterogeneous
analyte
, providing separate concentration and molecular weight variables for
two analytes.
B.6
L1 liposome capture
The recommended method for liposome capture on Sensor Chip L1 uses one
conditioning cycle consisting of two injections of 40 mM octylglucoside, followed
by three start-up cycles using a dummy sample before the actual samples are
run.