5 Methods
5.10 Requirements for assay-specific evaluation
100
Biacore T100 Software Handbook BR-1006-48 Edition AE
System preparations are equivalent to those for wizard-based runs (see
Section 4.2.5). The temperature settings are taken from the first assay step in
the method and cannot be changed here: they are shown for information only.
5.9.5
Rack positions
See Section 4.2.6 for a description of the
Rack positions
dialog box.
5.9.6
Prepare Run Protocol
See Section 4.2.7 for a description of the
Prepare Run Protocol
dialog box. Save
the method at this step to include all settings including the Prepare Run protocol
in the saved method.
5.9.7
Starting the run
When the positions are finalized, the sample plate is prepared and loaded into
the instrument and a chip is docked, choose
Start Run
to start the run. You wil
be asked to specify a file name for the results.
5.10 Requirements for assay-specific evaluation
This section describes the requirements and recommendations if assay-specific
evaluation is to be applied to method-based runs.
5.10.1 Concentration analysis
See Chapter 8 for a description of concentration evaluation.
Using calibration
•
At least one assay step is required with purpose
Calibration
and one with
purpose
Sample
. An assay step with purpose
Control Sample
is also
required in order to create trend plots for control samples.
•
Assay steps
Calibration
,
Sample
and
Control Sample
must be connected
to a cycle type that includes one
Sample
command. The three assay steps
will normally be connected to the same cycle type.
•
Samples in the
Calibration
step must have concentrations specified in the
variable
Conc
. At least two different concentrations are required for linear
calibration curves and at least four for 4-parameter fitting.
•
Samples in the
Sample
step will normally not have specified
concentrations. If concentrations are specified, they will be ignored.
Summary of Contents for Biacore T100
Page 1: ...GE Healthcare Biacore T100 Software Handbook ...
Page 2: ......
Page 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Page 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Page 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Page 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 237: ......