16-16
16.2 Processing Solution Replenishment System
Removal
1. Remove the PSR pump bracket
.
2. Pinch the inlet and outlet hoses with the pinchcocks.
3. Loosen the hose clamps and disconnect the
hoses from the filter.
Installation
Installation is essentially in the reverse order of removal.
NOTE: • Connect the hoses securely
• After installation, perform the following adjustment.
1) Menu 0620 "Pump Output Measurement/
Setting"
2) Menu 0621 "Auto Cleaning Output
Measurement/Setting"
Removal
1. Remove the connector cover
.
2. Disconnect the sensor connectors.
3. Open the three clamps and cut the cord tie.
16.2.7
PSR Filter Replacement
LII713
Hoses (2)
Pinchcocks (2)
Clamp
Filter
16.2.8
PSR Upper/Lower Level Sensor (FS723/FS727) Replacement
LII719
Connector (FS723)
Connector (FS727)
Clamps (3)
Cord Tie