5-121
5.7 Processor Adjustment/Maintenance (06)
5
3) Remove the screw and then the splash-
prevention cover.
4. Click the [Start] button for the pump to be
measured.
• The message below appears on the printer
operation panel.
5. Press the [Enter] key on the printer operation
panel.
• The alarm beeps intermittently for ten seconds.
NOTE: Click the [Cancel] button to stop the operation.
6. Before the pump starts discharging, disconnect
the replenisher nozzle, insert it into the measuring
cylinder and hold it at the same height as its
setting position.
• The replenisher pump starts discharging.
• The alarm beeps when discharging is
completed.
7. Insert the replenisher nozzle into the sub-tank.
8. Place the measuring cylinder on a level surface,
and then read and record the output.
L
2
1051
Screw
Splash-prevention Cover
#PSR MEASURE
#START [YES]
?
L
2
1050
PSR Replenisher Nozzle
Measuring Cylinder