3-159
3.2 Messages and Actions
3
E-7172
Reading FlashROM data from carrier
failed.
Consult your technical
representative.
Abnormal carrier/FlashROM data
(Causes of the error message)
1. Poorly connected plug-in
connector
2. Faulty CYB23 circuit board
1. Remove and reinstall the carrier. If
the error occurs again, replace the
plug-in connector.
2. Replace the circuit board.
W-7180
The sensor for 135 feeding lane
cannot be calibrated correctly.
Remove the carrier, and then clean
the glass surface of the upper and
lower feeding lanes with a dry cotton
swab.
It is possible to use the IX240 lane.
CYA23 135 sensor calibration error
(Causes of the error message)
1. Dust or dirt adhered to 135 frame
sensor (D101 to D115) glass
2. Faulty 135 frame sensor (LEE/
DTF23/DTG23)
3. Poorly connected or broken 135
frame sensor harness
4. Faulty flexible cable
5. Faulty JNC23 circuit board
6. Faulty CYA23 circuit board
1. Clean the glass.
2. Replace the sensor circuit board.
3. Reconnect or replace the harness.
4. Reconnect or replace the cable.
5. Replace the circuit board.
6. Replace the circuit board.
W-7182
The sensor for IX240 feeding lane
cannot be calibrated correctly.
Remove the carrier, and then clean
the glass surface of the upper and
lower feeding lanes with a dry cotton
swab.
You may use the 135 lane.
CYA23 IX240 sensor calibration error
(Causes of the error message)
1. Dust or dirt adhered to IX240
frame sensor (D116 to D119)
glass
2. Faulty IX240 frame sensor
(LEE22, DTF23 or DTG23)
3. Poorly connected or broken IX240
frame sensor harness
4. Faulty flexible cable
5. Faulty JNC23 circuit board
6. Faulty CYA23 circuit board
1. Clean the glass.
2. Replace the sensor circuit board.
3. Reconnect or replace the harness.
4. Reconnect or replace the cable.
5. Replace the circuit board.
6. Replace the circuit board.
W-7183
Film remains in IX240 feeding lane.
Remove carrier and then film.
Carrier installation error (IX240 film
remains)(unexposed film)
(Causes of the error message)
1. Dust or dirt adhered to IX240
unexposed frame sensor (D118/
D119) glass
2. Faulty IX240 unexposed frame
sensor
3. Poorly connected or broken IX240
unexposed frame sensor harness
4. Faulty flexible cable
5. Faulty JNC23 circuit board
6. Faulty CYA23 circuit board
1. Clean the glass.
2. Replace the sensor.
3. Reconnect or replace the harness.
4. Reconnect or replace the cable.
5. Replace the circuit board.
6. Replace the circuit board.
W-7184
The cartridge cannot be used.
Remove cartridge.
Under retracting operation (cartridge
is used when stopping cartridge
control function due to faulty cartridge
section).
(Causes of the error message)
1. Changed to degeneracy operation
due to faulty supply motor (D102),
door motor (M104), door open/
close sensor (D120) or supply
motor home position sensor
(D122)
2. Changed to degeneracy operation
due to faulty supply motor (D104),
door motor (M104), door open/
close sensor (D120) or supply
motor home position sensor
(D122) harness
3. Changed to degeneracy operation
due to faulty CYA23 circuit board
1. Replace the damaged parts.
2. Reconnect or replace the harness.
3. Replace the circuit board.
No.
Message
Factor
Actions