3-66
3.2 Messages and Actions
E-4274
An error occurred in the image
processing section.
Check the print is correctly done.
JPEG parameter abnormal (Script
parameter)
(Causes of the error message)
1. DIC setup (Quality Factor)
abnormal
2. Poorly connected or broken 1394
interface cable
3. Poorly inserted GIE23 circuit
board
4. Faulty GIE23 circuit board
1. Change the DIC setup (Quality
Factor).
2. Connect the cable correctly or
replace it.
3. Insert the circuit board correctly in
the slot.
4. Replace the circuit board.
E-4275
The image processing is not
operable.
Consult your technical
representative.
Processed plural DSP at DIMM
512MB
(Causes of the error message)
1. Large-size or high-magnification
print was printed when DIMM was
degenerated with 512MB.
2. Degenerated condition of DIMM
1. Do not print large-size or high-
magnification prints.
2. Reconnect or replace the DIMM.
W-4311
Carrier is not installed.
Install the carrier.
Carrier not installed.
(Causes of the error message)
1. Initializing or removed carrier
2. Communication error between
main control unit, GMB circuit
board and carrier
1. Execute again.
2. Check and correct connection.
I-4312
Correcting the scanner. Please wait a
moment.
It takes about 2 minutes.
Under scanner correction
Wait until completion
I-4313
Downloading the scanner correction
data.
Please wait.
Downloading
Wait until completion.
I-4314
The scanner temperature control is
not completed.
Please wait.
After the completion of temperature
control, the selected menu will be
automatically started.
Under temperature control
Wait until completion.
W-4315
The correction table is not found.
Perform the scanner correction in the
pre-operational check.
No brightness correction table for the
installed carrier when carrying out the
scanner correction.
Perform the scanner correction in the
pre-operational check.
W-4316
The mask is installed in the carrier.
Couldn’t perform the scanner
correction. Remove the mask.
MFC carrier installed with mask when
carrying out scanner correction (pre-
operational checks)
(Cause of the error message)
1. Not installed or seated calibration
mask
2. Faulty MSB23 circuit board
1. Install the calibration mask
properly.
2. Inspect the mask ID sensor by
input check and replace the circuit
board if it is abnormal.
W-4317
The sliding stage is not inserted.
Insert it, and then perform the
scanner correction with calibration
mask.
Slider of MFC carrier is not set when
carrying out scanner correction (pre-
operational checks).
(Causes of the error message)
1. Mask table slider out of optical axis
side lock position
2. Faulty mask table slider position
sensor (D486)
1. Lock the mask table slider in the
optical axis side.
2. Replace the sensor.