3-134
3.2 Messages and Actions
W-7028
Cartridge remains in carrier.
Remove the carrier, and then take out
the cartridge.
IX240 cartridge existing carrier
installation error
(Causes of the error message)
1. Installed carrier or turned power
ON with unexposed cartridge
remaining in carrier
2. Inserted unexposed cartridge
immediately after carrier installed
or power ON
3. Cartridge remains when Setup
and Maintenance mode is
selected.
4. Faulty cartridge set sensor (D123)
or harness
5. Faulty IPI sensor (D121) or
harness
6. Faulty CYA23 circuit board
1. Do not insert an unexposed
cartridge.
2. Do not insert an unexposed
cartridge.
3. Remove the cartridge and reinstall
the carrier.
4. Reconnect or replace the harness
or sensor.
5. Reconnect or replace the harness
or sensor.
6. Replace the circuit board.
W-7029
Feeding of film failed.
Remove carrier and then film.
MTP center position detected at fine-
scan is differ by 5mm or more from
one at pre-scan.
(Causes of the error message)
1. Slipped or blocked film
2. Broken cartridge door/flange
3. Faulty nest section unit
4. Faulty IX240 winding unit
5. Dust or dirt adhered to IX240
perforation sensor (D117L/P)
glass
6. Faulty IX240 perforation sensor
(D117L/P) or harness
7. Faulty feed/supply/door motor
(M101/M102/M103)
8. Faulty IX240/135 side belt
9. Faulty harness
10.Faulty flexible cable
11.Faulty JNC23 circuit board
12.Faulty CYA23 circuit board
1. Clean the feed rollers or clear the
cause of blocking.
2. If the error occurs with a certain
cartridge, reload film into a known
good cartridge.
3. Replace the nest section unit.
4. Replace the IX240 winding unit.
5. Clean the glass.
6. Replace the sensor or harness.
7. Replace the motor.
8. Adjust or replace the belt.
9. Reconnect or replace the harness.
10.Reconnect or replace the cable.
11.Replace the circuit board.
12.Replace the circuit board.
W-7030
Feeding of film failed.
Remove carrier and then film.
MTP hole length is more than 8mm at
time of feeding, or ATP cannot be
found after feeding 29mm from MTP.
(Causes of the error message)
1. Slipped or blocked film
2. Broken cartridge door/flange
3. Faulty nest section unit
4. Faulty IX240 winding unit
5. Dust or dirt adhered to IX240
perforation sensor (D117L/P)
glass
6. Faulty IX240 perforation sensor
(D117L/P) or harness
7. Faulty feed/supply/door motor
(M101/M102/M103)
8. Faulty IX240/135 side belt
9. Faulty harness
10.Faulty flexible cable
11.Faulty JNC23 circuit board
12.Faulty CYA23 circuit board
1. Clean the feed rollers or clear the
cause of blocking.
2. If the error occurs with a certain
cartridge, reload film into a known
good cartridge.
3. Replace the nest section unit.
4. Replace the IX240 winding unit.
5. Clean the glass.
6. Replace the sensor or harness.
7. Replace the motor.
8. Adjust or replace the belt.
9. Reconnect or replace the harness.
10.Reconnect or replace the cable.
11.Replace the circuit board.
12.Replace the circuit board.
No.
Message
Factor
Actions
ATP = Anticipation perforation MTP = Metering perforation