5-53
5.4 Scanner Adjustment/Maintenance (03)
5
3. Add 9900 pulses to the pulse value in the “Conj.
length” box and enter it in the “Conj. length” box.
NOTE: The added value may be approximately 11000.
If it exceeds 11000, enter 11000 in the “Conj.
length” box.
4. Click the [Move] button.
5. Turn ON the diffusion box sensor.
6. Click the [Update] button.
• The real-time image of the LED is displayed.
7. Click the right-hand button of the mouse while
pressing the [Shift] key to reduce the image.
8. Check the condition of LED pattern.
LED Pattern
R: 8
×
8
G: 5
×
8
B: 5
×
10
IR: 12
×
12
NOTE: Do not sfare at the LED light source for
prolonged periods of time.
This procedure sets the lens ID to be used and focal
length.
[Selection]
[Setup and Maintenance]
→
[03 Scanner Adjustment/
Maintenance]
→
[0344 Lens Registration]
• The current status is displayed.
5.4.10
Lens Registration (0344)