3-98
3.2 Messages and Actions
NOTE 1: The refreshing backup can be restored
(See Section 3.4)
.
E-4611
The control for the CLE circuit board
is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-message delivery
semaphore return error
(Causes of the error message)
1. Faulty system software
2. Faulty CLE23 circuit board (CPU)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
E-4612
The control for the CLE circuit board
is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-message delivery
semaphore reference error
(Causes of the error message)
1. Faulty system software
2. Faulty CLE23 circuit board (CPU)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
E-4613
The control for the CLE circuit board
is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-system initialization error
(Causes of the error message)
1. Faulty system software
2. Faulty CLE23 circuit board (CPU)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
E-4614
The control for the CLE circuit board
is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-task creation error
(Causes of the error message)
1. Faulty system software
2. Faulty CLE23 circuit board (CPU)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
E-4615
The control for the CLE circuit board
is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-mail box creation error
(Causes of the error message)
1. Faulty system software
2. Faulty CLE23 circuit board (CPU)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
E-4616
The control for the CLE circuit board
is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-alarm creation error
(Causes of the error message)
1. Faulty system software
2. Faulty CLE23 circuit board (CPU)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
E-4617
The control for the CLE circuit board
is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-fixed length memory pool
creation error
(Causes of the error message)
1. Faulty system software
2. Faulty CLE23 circuit board (CPU)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
E-4618
The control for the CLE circuit board
is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-variable length memory pool
creation error
(Causes of the error message)
1. Faulty system software
2. Faulty CLE23 circuit board (CPU)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
E-4619
The control for the CLE circuit board
is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-event creation error
(Causes of the error message)
1. Faulty system software
2. Faulty CLE23 circuit board (CPU)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
No.