3-119
3.2 Messages and Actions
3
W-4805
The optical magnification calibration
is not performed yet.
Perform it.
Optical magnification calibration was
not performed at the time of
executing focus calibration.
Perform the optical magnification
calibration.
W-4806
The focus calibration is not performed
yet.
Perform it.
Focus calibration was not performed
at the time of executing focus position
adjustment.
Perform the focus calibration.
I-4807
Downloading the bright correction
table.
Please wait.
Brightness cancel button was clicked. Wait until downloading is completed.
W-4808
Downloading of the bright correction
table failed.
Brightness correction table was not
found after brightness cancel button
was clicked.
(Causes of the error message)
1. IEEE1394 interface of GIE23
circuit board or main control unit is
unstable.
2. Faulty application on GMB23
circuit board
3. Faulty 1394 cable from image
processing box to main control unit
1. Restart the system software. If the
error occurs again, replace the
GIE23 or GMB23 circuit board.
2. Restart the system software. If the
error occurs again, replace the
GIE23 or GMB23 circuit board.
3. Reconnect or replace the cable.
W-4809
The feeding lane is not set to 135 in
the carrier.
Select the 135 lane.
NC100AY carrier was not in 135 feed
lane when executing in 135 feed lane.
(Causes of the error message)
1. Feed lane is not 135.
2. Faulty 135 feed lane position
sensor(D129)
3. Faulty 135 feed lane position
sensor(D129) harness
4. Faulty CYA23 circuit board
1. Set the carrier in the 135 feed
lane.
2. Replace the sensor.
3. Reconnect or replace the harness.
4. Replace the circuit board.
W-4810
The film mask type is incorrect.
Set the 135Full mask.
135F mask is not installed in
MFC10AY.
Install the 135F mask.
W-4811
Creating of the table failed.
Focus calibration table creation
failure
(Causes of the error message)
1. Incorrectly installed chart jig or
mask
2. Foreign matter in optical path
3. Abnormal mask pressure
1. Install the chart jig or mask
correctly.
2. Remove the foreign matter.
3. Replace the damaged pressure
cover parts or the solenoid.
W-4812
The floppy disk is not inserted into
the floppy disk drive.
Insert the floppy disk.
At the time of spectral calibration FD
insertion
(Causes of the error message)
1. FD not inserted
2. Not spectral calibration FD
3. Faulty FD drive
4. Faulty spectral calibration FD
1. Insert the FD.
2. Insert the correct FD.
3. Replace the main control unit.
4. Replace the FD.
I-4813
Correct the scanner.
Install the carrier.
At the time of scanner correction
execution confirmation for spectral
calibration
Install the carrier.
I-4814
Set the next film in the carrier.
No: ####
Specifying mounted film for spectral
calibration
Set the specified film.
I-4815
Reading the image.
Please wait.
At the time of image scanning for
spectral calibration
Wait until reading is completed.
No.
Message
Factor
Actions