5.7 Processor Adjustment/Maintenance (06)
This procedure measures and corrects the
replenisher pump output.
Normally measure only PSR.
NOTE: The machine automatically measures and adjusts
the pump output for P1R, P2RA and P2RB.
When executing this menu to confirm pump operation
for the P1R, P2RA or P2RB, return the discharged
replenisher into the replenisher tank.
[Selection]
[Setup and Maintenance]
→
[06 Processor
Adjustment/Maintenance]
→
[0620 Pump Output
Measurement/Setting]
1. Prepare a measuring cylinder.
2. Perform the following steps for PSR.
1) Open the water supply port door.
2) Loosen the screw and remove the water
supply port inner cover.
3. Perform the following steps for P1R, P2RA or
P2RB.
1) Open the replenisher box door.
2) Remove the circulation filter tray.
5.7.1
Replenisher Pump Output Measurement/Setting (0620)
L
2
1055
Screw
Water Supply Port Inner Cover
L
2
232
Circulation Filter Tray