8062A
Instruction Manual
2-24
for it, select the desired resistance range, short the test leads together, and
press the REL button. The REL indicator will appear and the display will
read zero. The 8062A will automatically subtract the stored test lead
resistance from subsequent measurements. Other common applications for
relative reference include: offset nulling (dc and ac voltage or current), diode
and transistor matching (diode test), resistor matching (
Ω
) and voltage
deviation (ac and dc voltage).
Note
The use of the relative function with the autoranging k
Ω
ranges is
restricted to the autoranging k
Ω
ranges. If you take a reference
reading within the autoranging k
Ω
range and then use it as a
reference outside autoranging k
Ω
, or use a reference reading taken
outside autoranging k
Ω
within the autoranging k
Ω
, errors will
result. There is no restriction on the use of the relative function
with the fixed resistance ranges or with autoranging M
Ω
.
Summary of Contents for 8062A
Page 4: ......
Page 8: ...8062A Instruction Manual iv...
Page 10: ...8062A Instruction Manual vi...
Page 14: ...8062A Instruction Manual 1 2...
Page 24: ...8062A Instruction Manual 2 2...
Page 50: ...8062A Instruction Manual 2 28...
Page 52: ...8062A Instruction Manual 3 2...
Page 62: ...8062A Instruction Manual 3 12...
Page 64: ...8062A Instruction Manual 4 2...
Page 90: ...8062A Instruction Manual 4 28...
Page 92: ...8062A Instruction Manual 5 2...
Page 102: ...8062A Instruction Manual 5 12 8062A 4031 iv39c eps Figure 5 2 A1 Main PCB Assembly...
Page 106: ...8062A Instruction Manual 6 2...
Page 108: ...8062A Instruction Manual 6 4 dy55c eps Figure 6 1 Accessories...
Page 118: ...8062A Instruction Manual 7 2...
Page 122: ...8062A Instruction Manual 7 6...
Page 123: ...8062A Instruction Manual 7 7 8062A 1201 iu46c eps Figure 7 5 A1 Main PCB Schmatic Diagram...
Page 124: ...8062A Instruction Manual 7 8 8060A 1003 iu61f eps Figure 7 6 A3 RMS PCB Schmatic Diagram...