1586A
Users Manual
3-20
Resistance Channels
Refer to Table 3-5 to configure a resistance channel.
Table 3-5. Resistance Channel Configuration
1
2
3
4
5
6
hcn008.eps
Item Function
Resistance function selection.
Set up the channel for a 2-wire or 4-wire resistance measurement connection. See “Sense Input
Configuration” on page 3-6.
Set the range to: Auto, 100
Ω
, 1 k
Ω
, 10 k
Ω
, 100 k
Ω
, 1 M
Ω
or 10 M
Ω
. Ch001 can be set to 100 M
Ω
.
Input a custom alpha-numeric label to help identify the channel (optional).
Set an alarm for this channel (optional). See “HI and LO Channel Alarms” on page 3-31.
Apply Mx+B scaling to the measurement (optional). See “Mx+B Scaling” on page 3-30.