Error Messages and Troubleshooting
Error Messages
7
7-13
Table 7-1. Error Messages (cont.)
Error Code
Error Message
Causes and Solutions
Error 431
Invalid setpoint
Cause:
Setpoint that is specified in the command is not in
range 1 to 20. This could be because of a wrong setpoint in
the TRIGger:AUTO:POINts, TRIGger:AUTO:SPOint,
TRIGger:AUTO:TOLerance, TRIGger:AUTO:STABility,
TRIGger:AUTO:DWELl commands
Solution:
Check the parameter value and try again.
Error 432
Input module not
installed
Cause:
The Input Module is not installed.
Solution:
Install the Input Module and try again.
Error 433
Recording is
automatic
Cause:
Recorder is set to OFF while automatic recording is in
progress.
Solution:
Stop the scan or use the ABORt command.
Error 434
Cannot record while
idle
Cause:
Recorder is started, but a scan has not been started
yet.
Solution:
Use command INITiate to start scan.
Error 435
Only fixed CJC
allowed on channel
Cause:
Ch001 is being configured to thermocouple with
internal or external CJC, that is not supported on this channel.
Solution:
Set up the input on a different channel.
Error 436
Unavailable
operation
Cause:
The operation is not supported on specified channel.
Solution:
Check the parameter value and try again.
Error 439
DIO currently
committed to scan
Cause:
DIO cannot be used as output while it is included in
scan list.
Solution:
Configure the setpoints.
Error 437
Setpoints not set
Cause:
User tried to start an Automated Test without
setpoints.
Solution:
Use command *RST to reset to factory default, or
set the DIO channel to OFF.
Error 440
Not able to achieve
requested resolution
Cause:
The Instrument cannot achieve the requested
measurement resolution. Invalid resolution in the CONFigure
or MEASure? Command could have been set.
Solution:
Check the parameter value and try again.
Error 441
Cannot monitor math
channel
Cause:
The user tried to monitor a math channel. Math
channels cannot be monitored.
Solution:
See the math channel data in the scan data screen.
Error 501
Invalid keyword or
choice
Cause:
Invalid keyword is found in the command.
Solution:
Check the parameter value and try again.
Summary of Contents for 1586A
Page 6: ...1586A Users Manual iv Error Messages 7 3 Troubleshooting 7 20...
Page 8: ...1586A Users Manual vi...
Page 10: ...1586A Users Manual viii...
Page 34: ...1586A Users Manual 2 2...
Page 42: ...1586A Users Manual 2 10 2 3 4 1 hcn018 eps Figure 2 6 Relay Card Installation...
Page 44: ...1586A Users Manual 2 12 hcn032 eps Figure 2 7 DAQ STAQ Multiplexer Installation...
Page 46: ...1586A Users Manual 2 14...
Page 48: ...1586A Users Manual 3 2...
Page 82: ...1586A Users Manual 3 36...
Page 84: ...1586A Users Manual 4 2...
Page 108: ...1586A Users Manual 5 2...
Page 114: ...1586A Users Manual 6 2...
Page 118: ...1586A Users Manual 6 6...
Page 120: ...1586A Users Manual 7 2...