Input and Channel Configuration
Channel Configuration
3
3-23
PRT Channels
Refer to Table 3-8 to configure a Platinum Resistance Thermometer (PRT) channel.
ITS-90 type requires deviation function coefficients to be entered. Coefficient "A" should
be set to the value of “a7” or “a8” as shown on the PRT calibration certificate. Likewise,
"B" is for “b7” or “b8” and "C" is for “c7”. If a coefficient is not available, it must be set
to “0”.
Table 3-8. PRT Channel Setup
1
2
3
4
5
6
7
8
hcn011.eps
Item Function
PRT
function
selection.
Select the PRT type. Types available: ITS90, CVD, PT-385, and PT-392.
Set the PRT resistance value at 0 °C (CVD, PT-385, or PT-392) or 0.01 °C (ITS-90).
Set up the channel for a 2-wire, 3-wire, or 4-wire measurement connection. See “Sense Input
Configuration” on page 3-6.
Input a custom alpha-numeric label to help identify the channel (optional).
Set an alarm for this channel (optional). See “HI and LO Channel Alarms” on page 3-31.
Apply Mx+B scaling to the measurement (optional). See “Mx+B Scaling” on page 3-30.
Load a PRT from the Probe Library. See “Probe Library” on page 3-34.
Summary of Contents for 1586A
Page 6: ...1586A Users Manual iv Error Messages 7 3 Troubleshooting 7 20...
Page 8: ...1586A Users Manual vi...
Page 10: ...1586A Users Manual viii...
Page 34: ...1586A Users Manual 2 2...
Page 42: ...1586A Users Manual 2 10 2 3 4 1 hcn018 eps Figure 2 6 Relay Card Installation...
Page 44: ...1586A Users Manual 2 12 hcn032 eps Figure 2 7 DAQ STAQ Multiplexer Installation...
Page 46: ...1586A Users Manual 2 14...
Page 48: ...1586A Users Manual 3 2...
Page 82: ...1586A Users Manual 3 36...
Page 84: ...1586A Users Manual 4 2...
Page 108: ...1586A Users Manual 5 2...
Page 114: ...1586A Users Manual 6 2...
Page 118: ...1586A Users Manual 6 6...
Page 120: ...1586A Users Manual 7 2...