Maintenance and Care
Introduction
6
6-3
Introduction
This section supplies information on how to clean the Product, reset the memory, and
replace the fuse in the rear of the Product.
Clean the Product
To clean the Product, wipe the instrument with a cloth that is lightly dampened with
water or mild detergent. Do not use aromatic hydrocarbons, chlorinated solvents, or
methanol based fluids.
Replace the Fuse
The Product has a product fuse that protects from overcurrent. Each voltage selection
requires a specific fuse. See Table 6-1.
Warning
To prevent possible electrical shock, fire, or personal injury,
use only specified replacement parts.
Table 6-1. Fuses
Voltage Selector
Fuse
Fluke Part Number
100 V
0.25 A, 250 V (slow blow)
166306
120 V
0.25 A, 250 V (slow blow)
166306
220 V
0.125 A, 250 V (slow blow)
166488
240 V
0.125 A, 250 V (slow blow)
166488
To replace the fuses (see Figure 6-1):
1.
Disconnect the mains-power cord from the power-entry module.
2.
Open the power entry module and remove the fuse holder.
3.
Replace the fuses with exact replacements as listed in Table 6-1.
Summary of Contents for 1586A
Page 6: ...1586A Users Manual iv Error Messages 7 3 Troubleshooting 7 20...
Page 8: ...1586A Users Manual vi...
Page 10: ...1586A Users Manual viii...
Page 34: ...1586A Users Manual 2 2...
Page 42: ...1586A Users Manual 2 10 2 3 4 1 hcn018 eps Figure 2 6 Relay Card Installation...
Page 44: ...1586A Users Manual 2 12 hcn032 eps Figure 2 7 DAQ STAQ Multiplexer Installation...
Page 46: ...1586A Users Manual 2 14...
Page 48: ...1586A Users Manual 3 2...
Page 82: ...1586A Users Manual 3 36...
Page 84: ...1586A Users Manual 4 2...
Page 108: ...1586A Users Manual 5 2...
Page 114: ...1586A Users Manual 6 2...
Page 118: ...1586A Users Manual 6 6...
Page 120: ...1586A Users Manual 7 2...