DC Voltage
Maximum Input
...................................................... 50 V on any range
Common Mode Rejection
..................................... 140 dB at 50 Hz or 60 Hz (1 k
Ω
unbalance in LOW lead) ±50 V peak
maximum
Normal Mode Rejection
........................................ 55 dB for power line frequency ±0.1 %, ±120 % of range peak
maximum
A/D Linearity
.......................................................... 2 ppm of measu 1 ppm of range
Input Bias Current
................................................. 30 pA at 25 °C
DC Voltage Accuracy
Accuracy is given as ± (% measu % of range). Basic accuracy specification is for Channel 1, medium or slow
sample rate. For channels x01 through x20 or when using Fast sample rate, add the numbers given in the table to the
accuracy specification. If the environment temperature is outside the specified range, multiply the temperature coefficient
numbers by the temperature deviation and add to the accuracy specification.
Range
Accuracy
Channel 1
Ch. x01 – x20
Fast
Sample Rate
T.C./ °C Outside
18 °C to 28 °C
±100 mV
0.0037 % + 0.0035 %
add 2
μ
V
add 0.0008 % of range
0.0005 % + 0.0005 %
±1 V
0.0025 % + 0.0007 %
add 2
μ
V
add 0.0008 % of range
0.0005 % + 0.0001 %
±10 V
0.0024 % + 0.0005 %
−
add 0.0008 % of range
0.0005 % + 0.0001 %
±50 V
0.0038 % + 0.0012 %
−
add 0.0008 % of range
0.0005 % + 0.0001 %
DC Voltage Input Characteristics
Range
Resolution
Input Impedance
Slow / Medium
Fast
±100 mV
0.1
μ
V 1
μ
V
10 G
Ω
[1]
±1 V
1
μ
V 10
μ
V
10 G
Ω
[1]
±10 V
10
μ
V 100
μ
V
10 G
Ω
[1]
±50 V
100
μ
V
1 mV
10 M
Ω
±1 %
[1] - Input beyond ±12 V is clamped. The clamp current is up to 3 mA.
DC Current
Input Protection
.................................................... 0.15 A resettable PTC
DC Current Accuracy
Accuracy is given as ± (% measu % of range). Basic accuracy specification is for medium or slow sample rate.
When using a fast sample rate, add the number given in the table to the accuracy specification. If the environment
temperature is outside the specified range, multiply the temperature coefficient numbers by the temperature deviation and
add to the accuracy specification.
Range Accuracy
Fast
Sample
Rate
T.C./ °C Outside
18 °C to 28 °C
±100
μ
A
0.015 % + 0.0035 %
add 0.0008 % of range
0.002 % + 0.001 %
±1 mA
0.015 % + 0.0011 %
add 0.0008 % of range
0.002 % + 0.001 %
±10 mA
0.015 % + 0.0035 %
add 0.0008 % of range
0.002 % + 0.001 %
±100 mA
0.015 % + 0.0035 %
add 0.0008 % of range
0.002 % + 0.001 %
Summary of Contents for 1586A
Page 6: ...1586A Users Manual iv Error Messages 7 3 Troubleshooting 7 20...
Page 8: ...1586A Users Manual vi...
Page 10: ...1586A Users Manual viii...
Page 34: ...1586A Users Manual 2 2...
Page 42: ...1586A Users Manual 2 10 2 3 4 1 hcn018 eps Figure 2 6 Relay Card Installation...
Page 44: ...1586A Users Manual 2 12 hcn032 eps Figure 2 7 DAQ STAQ Multiplexer Installation...
Page 46: ...1586A Users Manual 2 14...
Page 48: ...1586A Users Manual 3 2...
Page 82: ...1586A Users Manual 3 36...
Page 84: ...1586A Users Manual 4 2...
Page 108: ...1586A Users Manual 5 2...
Page 114: ...1586A Users Manual 6 2...
Page 118: ...1586A Users Manual 6 6...
Page 120: ...1586A Users Manual 7 2...