1586A
Users Manual
3-22
Thermistor Channels
Refer to Table 3-7 to configure a thermistor channel. The R(T) polynomial type requires
polynomial coefficients to be entered. The thermistor characterization polynomial is as
follows:
15
.
273
1
+
=
t
x
3
2
3
2
1
0
x
B
x
B
x
B
B
y
+
+
+
=
y
R
exp
=
Where t is the temperature in °C and R is the resistance in
Ω
. Any coefficients that are
not used must be set to 0.
Table 3-7. Thermistor Channel Setup
1
2
3
4
5
6
7
hcn010.eps
Item Function
Thermistor function selection.
Select the thermistor type. Types available: R(T), 2.252 k
Ω
, 5 k
Ω
, and 10 k
Ω
.
Set up the channel for a 2-wire measurement connection. See “Sense Input Configuration” on
page 3-6.
Input a custom alpha-numeric label to help identify the channel (optional).
Set an alarm for this channel (optional). See “HI and LO Channel Alarms” on page 3-31.
Apply Mx+B scaling to the measurement (optional). See “Mx+B Scaling” on page 3-30.
Load a thermistor from the Probe Library. See “Probe Library” on page 3-34.
Summary of Contents for 1586A
Page 6: ...1586A Users Manual iv Error Messages 7 3 Troubleshooting 7 20...
Page 8: ...1586A Users Manual vi...
Page 10: ...1586A Users Manual viii...
Page 34: ...1586A Users Manual 2 2...
Page 42: ...1586A Users Manual 2 10 2 3 4 1 hcn018 eps Figure 2 6 Relay Card Installation...
Page 44: ...1586A Users Manual 2 12 hcn032 eps Figure 2 7 DAQ STAQ Multiplexer Installation...
Page 46: ...1586A Users Manual 2 14...
Page 48: ...1586A Users Manual 3 2...
Page 82: ...1586A Users Manual 3 36...
Page 84: ...1586A Users Manual 4 2...
Page 108: ...1586A Users Manual 5 2...
Page 114: ...1586A Users Manual 6 2...
Page 118: ...1586A Users Manual 6 6...
Page 120: ...1586A Users Manual 7 2...