1586A
Users Manual
1-18
Thermocouple
Temperature Range
..............................................
−
270 °C to 2315 °C (depending on the sensor)
Voltage Range
.......................................................
−
15 mV to 100 mV
Thermocouple Voltage Accuracy
Accuracy is given as ± (|% of measurement| +
μ
V). Basic accuracy specification is for medium or slow sample rate. When
using a fast sample rate add the number given in the table to the accuracy specification. If the environment temperature is
outside the specified range, multiply the temperature coefficient numbers by the temperature deviation and add to the
accuracy specification.
Range
Accuracy
Channel 1
Ch. x01 – x20
Fast
Sample Rate
T.C./ °C Outside
18 °C to 28 °C
−
15 mV to 100 mV
0.004 % + 4
μ
V add
2
μ
V add
1
μ
V
0.0005 % + 0.0005 mV
Thermocouple Reference Junction Accuracy
Module CJC
Accuracy
T.C./ °C Outside
18 °C to 28 °C
DAQ-STAQ Module
0.25 °C
0.02 °C
High-Capacity Module
0.6 °C
0.05 °C
Thermocouple Temperature Accuracy
Accuracy specifications apply using medium or slow sample rate. When using fast sample rate, increase the accuracy
specification by 25 %. If the environment temperature is outside the specified range, increase the accuracy specification
by 12 % for every 1 °C outside the specified environment temperature range. Accuracy with fixed/external CJC does not
include the accuracy of the reference junction temperature. Linear interpolation may be used between points in the table.
Specifications do not include sensor accuracy. The practical range of temperature measurement depends on the sensor.
Type
(Range)
Temperature
Accuracy
Fixed / External CJC
Internal CJC
Channel 1
Ch. x01 – x20
DAQ-STAQ Module
High-Capacity
Module
K
−
270 °C to
1372 °C
−
200 °C
0 °C
1000 °C
0.28 °C
0.10 °C
0.14 °C
0.41 °C
0.15 °C
0.20 °C
0.76 °C
0.29 °C
0.32 °C
1.60 °C
0.62 °C
0.64 °C
T
−
270 °C to
400 °C
−
200 °C
0 °C
200 °C
400 °C
0.27 °C
0.10 °C
0.08 °C
0.08 °C
0.40 °C
0.15 °C
0.12 °C
0.11 °C
0.76 °C
0.30 °C
0.23 °C
0.20 °C
1.60 °C
0.65 °C
0.47 °C
0.41 °C
R
−
50 °C to
1768 °C
0 °C
300 °C
1200 °C
1600 °C
0.76 °C
0.42 °C
0.33 °C
0.34 °C
1.13 °C
0.63 °C
0.47 °C
0.49 °C
1.16 °C
0.64 °C
0.48 °C
0.50 °C
1.28 °C
0.71 °C
0.52 °C
0.54 °C
S
−
50 °C to
1768 °C
0 °C
300 °C
1200 °C
1600 °C
0.74 °C
0.45 °C
0.37 °C
0.39 °C
1.11 °C
0.67 °C
0.54 °C
0.56 °C
1.14 °C
0.68 °C
0.55 °C
0.57 °C
1.26 °C
0.76 °C
0.60 °C
0.63 °C
J
−
210 °C to
1200 °C
−
200 °C
0 °C
1000 °C
0.20 °C
0.08 °C
0.11 °C
0.29 °C
0.12 °C
0.14 °C
0.65 °C
0.28 °C
0.25 °C
1.41 °C
0.61 °C
0.53 °C
N
−
270 °C to
1300 °C
−
200 °C
0 °C
500 °C
1000 °C
0.42 °C
0.15 °C
0.12 °C
0.14 °C
0.62 °C
0.23 °C
0.17 °C
0.19 °C
0.90 °C
0.34 °C
0.24 °C
0.26 °C
1.69 °C
0.64 °C
0.44 °C
0.45 °C
Summary of Contents for 1586A
Page 6: ...1586A Users Manual iv Error Messages 7 3 Troubleshooting 7 20...
Page 8: ...1586A Users Manual vi...
Page 10: ...1586A Users Manual viii...
Page 34: ...1586A Users Manual 2 2...
Page 42: ...1586A Users Manual 2 10 2 3 4 1 hcn018 eps Figure 2 6 Relay Card Installation...
Page 44: ...1586A Users Manual 2 12 hcn032 eps Figure 2 7 DAQ STAQ Multiplexer Installation...
Page 46: ...1586A Users Manual 2 14...
Page 48: ...1586A Users Manual 3 2...
Page 82: ...1586A Users Manual 3 36...
Page 84: ...1586A Users Manual 4 2...
Page 108: ...1586A Users Manual 5 2...
Page 114: ...1586A Users Manual 6 2...
Page 118: ...1586A Users Manual 6 6...
Page 120: ...1586A Users Manual 7 2...