multi EA 4000
Transport and storage
143
8.
Remove the heating elements and the furnace cartridge from the multi EA 4000
and package these device components separately (
→
see section "Removing the
heating elements and furnace cartridge" p. 144).
1.
Switch off the C/S module from the mains switch.
2.
Disconnect the mains plug from the mains socket.
3.
Disconnect all connections on the rear of the module.
1.
Switch off the Cl module from the mains switch.
2.
Remove the measuring cell from the device and drain it (
→
see section "Servicing
3.
Rinse and dry the measuring cell and combined electrode.
4.
Unscrew and drain the adsorption tube.
5.
Remove and drain the halide trap.
6.
Re-install the empty adsorption tube and halide trap in the Cl module.
7.
Disconnect all connections on the rear of the module.
8.
Remove the sulfuric acid container and the safety container from the basic device
multi EA 4000. Drain and clean the sulfuric acid container (
→
see section "Replac-
ing the sulfuric acid/cleaning the sulfuric acid container" p. 114).
9.
Pull the measuring gas hose (transparent) and the bypass hose (black) out of the
channel in the multi EA 4000.
10.
Package the accessories carefully and in particular protect glass components
against breakage.
1.
Switch off the TIC solids module from the mains switch.
2.
Disconnect all connections on the rear of the module.
3.
Switch off the gas supply and disconnect the mains plug from the mains outlet.
4.
Put the acid container and drip tray aside. Drain and clean the acid container.
5.
Detach the hoses for gas and acid from the TIC reactor. Remove the TIC reactor.
Package the ceramic cannula separately.
6.
Drain the drying tube. Re-insert the drying tube.
7.
Drain the condensate container.
8.
Remove and clean the pump hose for acid from the cartridge pump.
9.
Carefully pull the condensation coil out of its mount.
10.
Package the accessories carefully and in particular protect glass components
against breakage.
C/S module
Cl module
Automatic TIC solids
module
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...