Operation
multi EA 4000
72
For C/S measurements a ceramic combustion tube is used. It is chamfered at both ends
and symmetrical. The combustion tube is pushed into the gas sluice and the dust trap
at the chamfered ends.
Figure 42 Ceramic tube with dust trap and gas sluice
1 Dust trap with cooling ribs
2 Chamfered ceramic tube
3 Gas sluice with gas connection
Check the electrical connections and hose connections according to section "Position-
ing the multi EA 4000 C/S" p. 49ff.
1.
Assemble the gas sluice. Ensure correct fit of the sealing
ring (see arrow).
Screw the base element and pressure ring together using
3 hexagon socket screws. Tighten the screws evenly in
turn.
2.
For carbon determination: Fill one end of the ceramic
tube approx. 2.5 cm thick with quartz wool. Push the
quartz wool approx. 2-4 cm deep into the tube.
NOTICE! Do not push the quartz wool too far into the
tube, because the wool clumps together at high temper-
atures.
The filled end of the ceramic tube points later towards
the dust trap.
3.
Push the ceramic tube into the gas sluice up to the stop.
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...