Commissioning
multi EA 4000
62
5.4
Connecting and using the autosampler
CAUTION
There is a risk of burns at the hot the combustion furnace! Allow the basic device to
cool down sufficiently before positioning and aligning the autosampler!
During autosampler operation, there is a risk of burns at the hot boat and ceramic
hook! After pulling out of the furnace, the boat and hook remain hot for a long time.
CAUTION
Risk of crushing! The autosampler arm, the gripper and the boat carousel move during
activation, initialization, adjustment and measurement operation. Keep an adequate
distance to avoid having your hands crushed.
NOTICE
Only adjust the device in the switched-off state. Otherwise, the sensitive electronics
may be damaged.
Make sure that the autosampler is parallel to the basic device and level with the work-
ing tube! If the sample boat is inserted into the furnace at an angle or not at the right
height relative to the working tube, the ceramic hook might break or the work tube be
damaged.
5.4.1
Installing and adjusting the autosampler FPG 48
Connect the sampler to the analyzer as follows:
1.
Place the autosampler to the right of the analyzer.
2.
Plug the cable on the low voltage side of the table power
supply into the connection at the rear of the autosampler.
Connect the power supply to the mains.
3.
Connect the data cable supplied to the "sampler" interface
at the rear of the sampler. Connect the other end of the
cable to the "sampler" interface at the rear of the module
with the control unit (C/S module or Cl module).
4.
Connect the boat sensor to the "aux" connection at the rear
of the sampler.
5.
Plug the grounding conductor into the connection at the
rear of the basic device.
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...