multi EA 4000
Function and setup
35
The connections are located at the rear of the automatic TIC solids module.
Figure 25 Connections at the automatic TIC solids module
1 Mains switch
2 Gas connection for oxygen at the TIC reactor
3 Power input module with mains switch and
fuse holder
4 Gas connection "COM module for connection to
the
C/S module
5 " valve module" connection for connection to the
C/S module
6 "valve HT" connection remains empty
7 "COM HT" connection for connection to the com-
bustion module HTS1500 of the multi EA 4000
8 "waste" connection for the waste hose to drain the
condensate
9 Change-over valve
The automatic TIC solids module is equipped with two mains switches. Because the
switch at the mains input module under the condensation coil is poorly accessible,
there is a second one at the front. Both mains switches must be on during operation.
To deactivate the device, it is sufficient to use the switch at the front.
Connections
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...