Operation
multi EA 4000
84
1.
Screw the PTFE tray and the 2 clamps for the
safety container to the basic device, if applicable.
2.
Fill the sulfuric acid container with 12-15 ml sulfu-
ric acid.
3.
Hold the filled sulfuric acid container above the
PTFE tray.
4.
Attach the gas outlet of the combustion tube to
the inlet of the sulfuric acid container using the
spherical joint connection (1).
5.
Secure the spherical joint connection with the
forked clamp (2). Tighten the knurled head screws
finger-tight to do so.
6.
Push the measuring gas hose (2, transparent)
onto the hose connection of the gas olive.
7.
Seal the sulfuric acid container with the gas olive
(1).