Operation
multi EA 4000
94
If the tube gets jammed, push it further without pres-
sure by slightly rotating it. Insert the tube firmly into
the seal of the gas sluice.
5.
Place the dust trap contact guard on the left-hand side
of the basic device.
6.
Fit the heat guard above the gas sluice.
7.
Unscrew the 2 spacers from the multi EA 4000.
8.
Position the automatic TIC solids module to the right
of the basic device. Move the TIC solids module as
close to the wall of the basic device that the 2 white
spacers at the TIC solids module touch the device wall.
9.
Check the electrical connections and hose connections
(
→
see section "Positioning the automatic TIC solids
1 Ceramic cannula with sealing ring
2 Quartz glass TIC reactor
3 Pump hose with connection to TIC reactor
4 hose pump
5 Knurled ring
1.
Slide the TIC reactor (2) into the coupling until it touches
the ceramic tube. The connections of the TIC reactor must
point up.
2.
Attach the TIC reactor to the coupling. To do so, carefully
tighten the knurled ring (5).
NOTICE! The reactor must be centered in the coupling.
If the TIC reactor cannot be inserted in this manner, un-
screw the knurled ring. Push the knurled ring and the 2
green sealing rings onto the TIC reactor. Then insert the
TIC reactor into the coupling and tighten.
3.
Place the ceramic cannula with sealing ring (1) onto the
port with the screw thread.
4.
Clamp the pump hose in the cartridge of the hose pump
(4).
The hose end with the connection to the reactor (3) must
point towards the reactor.
Connect the TIC reactor
and sampler
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...