Fault removal
multi EA 4000
134
The respective status of the detector modules is indicated by color (see above). The
following status displays are possible:
Device ready for measurement
Display
Description
OK (green, black)
Detector is ready for measurement
Device not ready for measurement - general
Display
Description
no detector displayed
(display empty)
no detector identified:
activate detector
Initialize
communication error (rot)
communication interrupted:
deactivate/activate the device
Initialize
no connection (red)
connection interrupted:
check connection cable
deactivate/activate the device
Initialize
Device not ready for measurement - C-NDIR / S-NDIR
Display (rot)
Description
warning, analog values
analog values outside range:
Read the values via menu item S
YSTEM
COMPONENT TEST
on the tab C-NDIR;
S-NDIR,
See also fault description of the C/S module p. 135
warm-up time
detector is not yet ready for operation:
wait until the warm-up time is complete after activation
(approx. 30 min)
Device not ready for measurement - CI-AMP
Display (rot)
Description
inactive
no cell identified:
use cell
Device not ready for measurement - CI-AMP
Display (green)
Description
interval titration
interval titration running:
possible to start measurement
End point routine
end point routine running:
wait until the end point routine has finished
In the area of the flow indication the current intake flow (A
CTUAL
) and the target flow
to be reached (N
OMINAL
) are displayed. The target flow is 100 l/h. The intake flow may
differ ± 10 l/h from the target flow. Adjust the intake flow at the rotameter "pump" at
the front panel of the basic device. The temperature display shows the current furnace
temperature. If the temperature display is red, the current furnace temperature does
not match the temperature set in the method. In this case, wait until the target tem-
perature has been reached or check the temperature configuration in the method.
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...