Measuring methods
multi EA 4000
42
To determine the inorganic carbon (TIC), acid is added to an aliquot of the sample.
This leads to the decomposition of the carbonates/hydrogen carbonates contained
therein. The CO
2
is purged and detected. No thermal oxidation takes place.
3 (
CO
3
)
2
-
+ 2 H
3
PO
4
3 H
2
O + 3 CO
2
+ 2 (PO
4
)
3-
(3)
The sample digestion takes place in a TIC solids module. The measuring gas is detected
in the C/S module.
4.3
TOC analysis
During the TOC analysis the total organic carbon content of a sample is detected. In
the multi EA 4000 the TOC content can be determined in two ways:
Differential method
Direct method
The differential method should be used when the sample contains easily purgeable or-
ganic substances as benzene, cyclohexane, chloroform, etc. The differential method
should not be used when the TIC content of the sample is significantly higher than the
TOC content.
NOTICE
For the TOC analysis with the automatic TIC solids module the differential method is
preferred. The combustion tube and detectors are not exposed to acid in this case.
4.3.1
Differential method for the TOC analysis
The differential method can take place using the automatic TIC solids module. The
sample is weight in on two boats. Two consecutive measurements are used to deter-
mine TIC and TC. The calculated difference is given as TOC. The differential method
detects volatile as well as non-volatile organic carbon compounds.
TOC = TC – TIC
(4)
TOC
total organic carbon
TC
total carbon
TIC
total inorganic carbon
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...