multi EA 4000
Function and setup
25
Figure 11 Indications and connections at the C/S module
1 "PC" USB port for connection to the PC
2 "valve" connection – connection to the
flame sensor or automatic
TIC solids module (optional)
3 "COM" connection – connection to the
combustion module HTS1500 of the
multi EA 4000 or automatic TIC solids module
(optional)
4 "scale" connection – connection of the scales
5 "sampler" connection – connection of the
autosampler FPG 48
6 Cl-Coul – connection to the Cl module
7 Gas outlet "pump" – connection to the multi EA
4000
8 Gas connection "Cl" – connection to the Cl module
9 Gas inlet "analyte" – measuring gas inlet
10 mains plug connection
11 Fuse holder
12 Mains switch
3.4
Cl module
The Cl module consists of the following main components:
Wide-range coulometer for amperometry
Stirrer/cooling block for the measuring cells
Measuring cells with electrodes
Sulfuric acid container, with split outlet and safety container where applicable
Split valve
Buffer container as pressure compensation to compensate for the switching pulses
of the split valve
Adsorption tube
Halide trap
For the multi EA 4000 Cl the control electronics is also housed in the Cl module.
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...