Commissioning
multi EA 4000
56
5.3.4
Positioning the automatic TIC solids module
For the combination multi EA 4000 - C/S module - Cl module, the following connec-
tions must be in place:
multi EA 4000 connections
Target
"analyte out"
TIC module "analyte (HT)"
"O
2
"
Oxygen gas connection
Mains cable
Mains connection
"com1"
TIC module "COM (HT)"
"pump in"
C/S module "pump"
"Ar"
Gas connection for argon (only required for Cl
measurement and pyrolysis)
"waste out"
Open outlet for the exhaust air of the analyzer
Additional C/S module con-
nections
Target
"PC"
USB port for connection to the computer
"valve"
TIC module "valve (module)"
"scale"
External scales (optional)
"COM"
TIC module "COM (module)"
"sampler"
Connection of the autosampler FPG 48
"analyte"
TIC module "analyte (module)"
Socket for mains plug
Mains connection
Additional TIC module con-
nections
Target
"waste"
Connection for the waste hose to drain the conden-
sate
Socket for mains plug
Mains connection
Connections
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...