multi EA 4000
Fault removal
129
14
Data transfer interrupted
Cause
Remedy
no data transfer for 10 s
Initialize
17
Incorrect interface protocol ID
Cause
Remedy
Fault after update
(program versions of firmware multiWin don't match)
update required
20
Timeout: InitEnd
Cause
Remedy
timeout during initialization
Initialize
21
Timeout: StatusBusy
Cause
Remedy
Timeout during operation
(device not ready to measure)
Acknowledge message
Initialize
22
Timeout: End
Cause
Remedy
Timeout when exiting multiWin
Acknowledge message
Initialize
23
Timeout: StopEnd
Cause
Remedy
Timeout during measurement cancellation
Acknowledge message
Initialize
24
Timeout: Measurement
Cause
Remedy
Timeout during measurement
Maximum integration time has been exceeded
Check analyzer
Check method configuration
25
Timeout: Analyzer command processing
Cause
Remedy
Timeout during command processing
Acknowledge message
Initialize
30
Chloride needs adding
Cause
Remedy
Measuring cell and/or electrolyte solution contam-
inated
Combined electrode faulty or aged strongly
Clean measuring cell, fill fresh electrolyte
Use new electrode for chloride determination.
50
Firmware reset
Cause
Remedy
internal computer (firmware) restarted
Acknowledge message
Initialize
52
Firmware standby
Cause
Remedy
Firmware has been moved to standby
Acknowledge message. Initialize.
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...