Operation
multi EA 4000
106
Slowly adjust the stirrer to stage 5. The sample must not be spun against the
wall of the flask. Jolting or rapid movement of the stirrer must therefore always
be avoided.
Add 2 ml acid evenly 1-3 times (dependent on the concentration and sample
volume).
Then close the valve in the metering pump stop cock.
Since the release of CO
2
is also dependent on the metering speed and the stirring
speed of the stirrer magnet, always ensure the same cycle and the same speeds are
used in each measurement.
The analyzer starts processing the analysis sequence. The M
EASUREMENT
–
A
NALYSIS
window opens.
19.
After completing the measurement stop the magnetic stirrer and open the flask:
Detach the TIC head with a twisting motion. Move the arm of the stand-up, swivel
it back and lock it in this position.
20.
Clean and dry the flask.
21.
Connect a dry Erlenmeyer flask to the TIC head during breaks in measurements.
This maintains the oxygen flush (conditioning) and the dry gas flow frees the hose
system of moisture residue.
22.
Keep the heating plate at a constant temperature. Only switch off the thermostat
during longer measuring breaks.
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...