Transport and storage
multi EA 4000
142
9
Transport and storage
9.1
Transport
9.1.1
Preparing the analyzer for transport
WARNING
Risk of damage to health due to improper decontamination!
Perform a professional and documented decontamination of the device before return-
ing it to Analytik Jena. The decontamination report is available from Service when reg-
istering the return. Analytik Jena must refuse acceptance of contaminated devices. The
sender may be liable for any damage caused by inadequate decontamination of the
device.
CAUTION
Risk of burning at the hot furnace! Only remove the combustion furnace when the de-
vice is cold or allow the device to cool down sufficiently!
CAUTION
When removing the glass components there is a risk of injury from glass breakage! Re-
move all glass components carefully from the analyzer!
NOTICE
The heating elements and furnace cartridge can be damaged by vibration during
transport. Therefore, these device components must be removed from the multi EA
4000 and packaged separately before a transport. Reinstallation and recommissioning
of the basic device may only be performed by the Analytik Jena AG customer service or
trained specialist personnel authorized and trained by Analytik Jena AG.
Unsuitable packaging material and residue of measuring solution and chemicals can
damage individual components of the analyzer!
Only transport the analyzer in its original packaging! Ensure that the analyzer is fully
drained and all transport locks have been fitted!
Prepare the multi EA 4000 and its modules for transport as follows:
1.
Switch off the analyzer from the mains switch at the rear of the device and allow
the device to cool down.
2.
Switch off the gas supply and disconnect the mains plug from the mains outlet.
3.
Disconnect all connections on the rear of the analyzer.
4.
Open the doors of the analyzer. Drain the drying tube. Re-insert the drying tube.
5.
Unscrew the connections at the halide trap and press the halide trap out of its
clamps. Drain the halide trap.
6.
Pull the combustion tube out of the furnace.
7.
Package the accessories carefully and in particular protect glass components
against breakage.
Basic device
multi EA 4000
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...