Operation
multi EA 4000
102
pump
O
2
Ar/O
2
approx.1.7 l/min
corresponds to N
OMINAL
=
100 ± 10 in the S
TATUS ANA-
LYZER
window
2.5 l/min
1.5 l/min
The analysis system now has time to settle down.
If the intake flow differs greatly or changes later, readjust the intake flow at the
"pump" valve of the multi EA 4000.
8.
Weigh the sample on a ceramic boat.
9.
Start the measurement by clicking on [S
TART MEASUREMENT
].
10.
Select or create a new analysis group for storing the measuring data.
11.
Create an analysis sequence.
Add the desired number of samples to the sample table with
.
Enter the following data:
Column
Description
S
AMPLING POSITION
Position of the sample on the sample rack
The sample rack must only be populated with ascending po-
sition numbers. However, it does not need to be without
gaps.
N
AME
Sample ID
With
an automatic sample ID can be created.
A
NALYSIS TYPE
Selection of the analysis type (S
AMPLE
,
D
AILY FACTOR DETERMI-
NATION
,
C
ALIBRATION MEASUREMENT
,
AQA
MEASUREMENT
or
D
ET
.
BOAT BLANK
)
N
OMINAL VALUE
Target concentration for calibration standards or AQS
standard
O
PERANDS
Input of the sample weight
Populate the autosampler. Use the
and
buttons to move the sample
rack to the desired position.
Click on in the sequence row to release it for analysis. All sequence rows can
be released by clicking on the
button.
Sequence rows released for analysis have a green fill.
12.
Click on [S
TART MEASUREMENT
].
The analyzer starts processing the analysis sequence. The M
EASUREMENT
–
A
NALYSIS
window opens.
The boat pick-up picks up a boat from the sample rack and places it onto the
ceramic hook.
The ceramic hook moves the boat into the TIC reactor. As soon as the boat has
reached the position under the ceramic cannula, the acid addition starts auto-
matically. During the slow boat feed acid is added to the entire sample.
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...