Fault removal
multi EA 4000
130
61
62
64
65
66
67
Command from the PC not complete
Command from the PC without STX
Command from the PC CRC error
Command from the PC invalid command
Command from the PC invalid MEAS command
Command MTXT from PC missing
Cause
Remedy
communication error
Acknowledge message
Initialize
100
C sensor: no connection
Cause
Remedy
Communication faulty after sensor
was detected during initialization
Acknowledge message
Initialize
101
C detector: CRC error
Cause
Remedy
Communication faulty after sensor
was detected during initialization
Acknowledge message
Initialize
104
107
C sensor: Analog values out of range
C sensor: Analog values out of range (coarse)
Cause
Remedy
The analog values of the detector
are outside the operating range
Check the quality of the carrier gas
Initialize
Check the analog values in the component test (via
menu item S
YSTEM
C
OMPONENT TEST
tab C-NDIR)
106
C sensor: incorrect response to command
Cause
Remedy
communication faulty
Switch C module off and back on
Initialize
120
130
S sensor: no connection
Cl sensor: no connection
Cause
Remedy
Communication faulty after sensor
was detected during initialization
Acknowledge message
Initialize
120
S sensor: no connection
Cause
Remedy
Communication faulty after detector
was detected during initialization
Acknowledge message
Initialize
124
S sensor: incorrect status
Cause
Remedy
Change of smoothing factor
Repeat action
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...