multi EA 4000
Operation
73
4.
Insert the ceramic tube with gas sluice into the multi EA
4000 on the right-hand side. If the tube gets jammed,
push it further without pressure by slightly rotating it.
5.
Screw the gas sluice with 3 hexagon socket screws to the
basic device. The gas connection must point to the front
of the basic device.
6.
Attach the oxygen hose (hose no. 17) to the gas sluice.
To do so, push the hose onto the gas connection and se-
cure it with the union nut.
7.
Pull off the dust trap contact guard from the opening on
the left side of the basic device (see Figure below).
8.
Place the dust trap onto the ceramic tube. Align the dust
trap with the gas connection pointing up.
Slide the dust trap towards the tube axis onto the tube
with a rotating motion and push slightly until the dust
trap slides onto the tube up to the stop.
9.
To do so, push the measuring gas hose (hose no. 1) onto
the gas connection and secure it with the union nut.
10.
Replace the dust trap contact guard.
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...