Function and setup
multi EA 4000
40
3.7
Additional optional accessories
3.7.1
Sampler
Two versions of sampler are available for the sample feed of the multi EA 4000:
the solid autosampler FPG 48
the manual feed
The autosampler FPG 48 offers space for 48 sample boats. The boats are automatically
picked from the trays and transferred to the combustion furnace. The feed for the
transfer is computer-controlled. Dependent on the equipment the boats are placed
back on the autosampler after the measurement or discarded in a waste container af-
ter cooling.
For analyses in the quartz combustion tube (Cl measurements) the FPG 48 can be
combined with a flame sensor.
The manual feed is a robust device facilitating the movement of the ceramic hook with
the straight aligned sample boats into the combustion furnace. In the manual sample
feed it is important that with highly reactive samples the transfer into the combustion
tube must be slow and controlled.
3.7.2
Flame sensor
A flame sensor can optionally be combined with the autosampler FPG 48 for Cl deter-
mination. The flame sensor captures the combustion of highly reactive samples in the
quartz tube visually and controls the boat advance in accordance with the brightness
of the flame. The ensuing optimized combustion characteristics lead to precise meas-
uring results.
3.7.3
Scales
Scales can be connected for the automatic volume transfer for the calculation of the
analysis results. The samples must be weighted in accordance with the instructions in
the program. The determined weights are automatically transferred to the sample ta-
ble.
Autosampler FPG 48
Manual feed
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...