Maintenance and care
multi EA 4000
124
1.
Push the bracket on the condensate
pump to the left.
2.
Pull the hoses no. 77 and no. 78 off the
connections.
3.
Remove the conveyor belt with the pump hose from the pump body.
4.
Check the pump hose and the connections on excessive wear and cracks.
NOTICE! If moisture escapes from the pump hose or the connections, the pump
hose must be replaced.
5.
Wipe the pump body and roller carrier with ultrapure water.
6.
Check the pump body and roller carrier for wear.
NOTICE!
Notify customer service if pump body and roller carrier are strongly af-
fected.
1 Conveyor belt
2 Groove
3 Metal adapter
4 Hose guide
5 Hose clamp
6 Pump hose
7.
Push the intact or new pump hose back
into the conveyor belt.
During installation, the hose clamps
must be rotated downwards.
Push the hose guide into the groove on
the conveyor belt.
8.
Position the conveyor belt around the
pump body.
9.
Press the conveyor belt upwards with
one hand and rotate the bracket with
the other hand to the right until it en-
gages.
10.
Push hoses 77 and 78 back onto their
adapters.
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...