Transport and storage
multi EA 4000
144
9.1.2
Removing the heating elements and furnace cartridge
WARNING
Life-threatening electrical voltages and high currents may occur in the interior of the
analyzer! For the removal of the heating elements and furnace cartridge witch off the
analyzer from the mains switch at the rear of the device and disconnect the mains plug
from the mains outlet.
The heating elements and furnace cartridge may only be reinstalled by the customer
service department of Analytik Jena AG or by specialist personnel authorized and
trained by Analytik Jen. AG! The recommissioning of the basic device must also be per-
formed by these persons.
CAUTION
There is a risk of burns on the furnace! Only do the removal when the device is cold or
allow the device to cool down sufficiently (best overnight)!
Caution when opening the right side panel. There is a risk of injury from sharp-edged
components.
The high temperature ceramic heating elements and furnace cartridge can be dam-
aged by vibration during transport. Therefore, the device components must be re-
moved from the basic device and packaged separately before a transport. The heating
elements and furnace cartridge need not be removed for moving within the laboratory.
1.
If not yet done, remove the dust trap through the open-
ing in the left side panel of the basic device.
2.
Unscrew the gas sluice. Remove the combustion tube
from the furnace.
3.
Open the right side panel of the basic device. To do so,
undo the 4 hexagon socket screws. Do not unscrew the
screws fully.
4.
Detach the protective conductor (1) and the two gas
hoses (2,3, hose no. 13 and 16) from the inside of the
side panel. Remove the side panel.
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...