9-8
129373-496
Operating Instructions
Programming Operating Parameters
5. Once the correct "too long in phase" values have been entered, press
RETURN
on screen #53. Control saves all changes made and display
returns to screen #14.
Pressing PREVIOUS returns display to
screen #40.
Pressing MAIN MENU returns display to
screen #1.
53
CHANGE TOO LONG IN PHASE PARAMETERS
TOO LONG IN:
AIR BREAK = 30:00
DOOR SEAL = 30:00
DOOR UNSEAL = 30:00
JACKET CHARGE = 30:00
PREVIOUS
MAIN
MENU
RETURN
Sterilizer uses the setup options to assign the temperature and pressure units,
control volume of audible signals, control utilities sterilizer shutdown feature,
and control printer functions.
1. Access Change Values menu (screen #14) as described at the beginning
of this section.
2. Press
SETUP
.
9.4 Setup
Pressing RETURN returns display to
screen #13.
Pressing MAIN MENU returns display to
screen #1.
14
CHANGE VALUES
RETURN
MAIN
MENU
TOO LONG
IN PHASE
TIME/
DATE
SETUP
CYCLES