6-27
Sterilizer Operation
Operating Instructions
129373-496
4. To begin Dart Warm-up, press
DART WARMUP
.
3. Press
TEST CYCLES
button.
1.
PREVAC
4.
PREVAC
2.
GRAVITY
5.
GRAVITY
3.
LIQUID
6.
LIQUID
2
CYCLE SELECT
Pressing MAIN MENU returns display to
screen #1.
MAIN
MENU
TEST
CYCLES
3
CYCLE SELECT
LEAK
TEST
DART
DART
TEST
WARMUP
Pressing STANDARD CYCLES returns
display to screen #2.
Pressing MAIN MENU returns display to
screen #1.
MAIN
MENU
STANDARD
CYCLES
NOTE: If
DART WARM-UP
is pressed while the chamber door is open, a
display screen will appear directing the operator to close door before
continuing cycle operation. Operator must reselect DART WARM-UP after
the door is closed.
NOTE: If a cycle is started when the sterilizer has not been fully calibrated,
a display screen will appear indicating that the control is not calibrated.
Sterilizer must be calibrated by a qualified service technician before process-
ing loads.
5. Sterilizer automatically progresses through the following cycle phase:
NOTE: If an alarm occurs during cycle operation, refer to
S
ECTION
11, R
OUTINE
M
AINTENANCE
, for instructions on correcting the alarm condition.
NOTE: If power is lost during cycle operation, cycle either continues in same
phase or aborts if seal pressure is below 5 psig (0.34 Pbar) once power is
restored.
• Jacket Charge
– Jacket charges with steam to 131°C (268°F).