12-20
129373-496
Operator Manual
Troubleshooting
Screen with Service Instructions
Screen with Operator Instructions
12.4.7 Jacket
Temperature
Probe Failure
Occurs if jacket tem-
perature reading is
outside the normal
range.
Occurs if door seal
switch contact is made
but door switch is still
open.
12.4.8 Door Switch
Failure
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
CHAMBER:
000.0 C
0.0 psig
JACKET:
000.0 C
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
STATUS..
ALARM! JACKET
213
TEMPERATURE PROBE FAILURE
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
237
DOOR SWITCH FAILURE
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
•
EXHAUST CHAMBER TO ATMOSPHERIC
PRESSURE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
Alarm
Description
Screen with Operator Instructions
STATUS..
SERVICE INFORMATION:
2 1 4
JACKET TEMP PROBE FAILURE
→
→
→
→
→
RTD PROBE, RTD3, OUTPUT IS OUTSIDE
NORMAL RANGE
CAUSES AND CORRECTION:
1.
LOOSE CONNECTION IN PROBE WIRING
→
→
→
→
→
REPAIR
2.
PROBE FAILED
→
→
→
→
→
REPLACE
→
→
→
→
→
RECALIBRATE
3.
CONTROL OUT OF CALIBRATION
→
→
→
→
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4.
MAIN CONTROL FAILURE
→
→
→
→
→
CHECK CONTROL BOARD POWER STATUS
LEDs
→
→
→
→
→
REPLACE CONTROL BOARD
→
→
→
→
→
RECALIBRATE
PAPER
SERVICE
EXIT
FEED