12-3
Troubleshooting
Operator Manual
129373-496
Occurs if chamber does
not exhaust to 4 psig
(0.28 bar) within the
allotted time.
Screen with Service Instructions
Screen with Operator Instructions
Occurs if chamber does
not reach the set
evacuation level within
the allotted time.
12.2.2 Too Long In
Exhaust
12.2.3 Too Long In
Evacuation
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
•
EXTEND EXHAUST TIME
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
STATUS..
ALARM!
200
TOO LONG IN EXHAUST
SILENCE
SERVICE
ALARM
HELP
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK WATER SUPPLY VALVE
→
→
→
→
→
IF CLOSED, OPEN VALVE
3. IF ALARM RECURS, ABORT CYCLE
AND CALL SERVICE
STATUS..
ALARM!
202
TOO LONG IN EVACUATION
ABORT
Alarm
Description
Screen with Operator Instructions
→
→
→
→
→
CHAMBER DID NOT EXHAUST TO
ATMOSPHERIC PRESSURE WITHIN ALLOTTED
TIME
CAUSES AND CORRECTION:
1.
CHAMBER DRAIN STRAINER PLUGGED
→
→
→
→
→
CLEAN
2.
SOLENOID VALVE MALFUNCTION
→
→
→
→
→
REPAIR S03
3.
CONTROL OUT OF CALIBRATION
→
→
→
→
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
STATUS..
SERVICE INFORMATION:
201
TOO LONG IN EXHAUST
PAPER
SERVICE
EXIT
FEED
MODE