5-5
Control Interface
Operating Instructions
129373-496
5.3 Out-of-cycle
Options Menus
All other sterilizer functions, including cycle programming and printer opera-
tion, are accessed through the Out-of-cycle Options menu screens (#13 and
#87).
S
ECTION
10, O
UT
-
OF
-
CYCLE
O
PTIONS
, describes each function accessible
from these menu screens.
After pressing OPTIONS on screen #1, screen #13 appears showing six out-
of-cycle functions.
•
Pressing
STATUS PRINT
generates a printout listing the time of day and
current readings from the pressure and temperature probes. Refer to
S
ECTION
10 for more information.
•
Pressing
DUPLICATE PRINT
generates a printout of cycle data from the
last completed cycle. Refer to
S
ECTION
10 for more information.
•
Pressing and holding
PAPER FEED
continually advances the printer paper.
Refer to
S
ECTION
9 for more information.
•
Pressing
CHANGE VALUES
provides access to the Change Values menu.
User may program the cycle values and sterilizer operating parameters
from the Change Values menu. Refer to
S
ECTION
8, P
ROGRAMMING
C
YCLE
V
ALUES
,
and
S
ECTION
9, P
ROGRAMMING
O
PERATING
P
ARAMETERS
, for further information.
•
Pressing
DISPLAY VALUES
allows user to view the current programmed
cycle values and operating parameters. Refer to
S
ECTION
10 for more
information.
•
Pressing
PRINT VALUES
allows user to generate a printout of the current
programmed cycle values and operating parameters. Refer to
S
ECTION
10 for
more information.
13
OUT-OF-CYCLE OPTIONS
STATUS
CHANGE
VALUES
DUPLICATE
DISPLAY
VALUES
PAPER
FEED
VALUES
NEXT
MAIN
MENU
Pressing NEXT advances display to
screen #87.
Pressing MAIN MENU returns display to
screen #1.