12-12
129373-496
Operator Manual
Troubleshooting
12.3.2
Too Long
To Open Door
Occurs if door switch
does not open within
the allotted time.
Screen with Operator Instructions
12.3.3 Pressure In
Chamber
Occurs if 2 psig
(0.14 Pbar) pressure
is sensed in the
chamber.
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
221
PRESSURE IN CHAMBER
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY EXHAUST CHAMBER TO
ATMOSPHERIC PRESSURE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
STATUS..
SERVICE INFORMATION:
246
TOO LONG TO OPEN DOOR
PAPER
CLEAR
EXIT
FEED
ALARM
→
→
→
→
→
DOOR SWITCH DID NOT OPEN IN
ALLOTTED TIME
CAUSES AND CORRECTION:
1.
DOOR SWITCH MALFUNCTION
→
→
→
→
→
CHECK LS1(LS2) CONNECTIONS
→
→
→
→
→
READJUST LS1(LS2)
2.
POWER DOOR MECHANISM FAILURE
→
→
→
→
→
REPAIR MECHANISM
→
→
→
→
→
REPLACE MOTOR
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
245
TOO LONG TO OPEN DOOR
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
REMAIN IN ALARM CONDITION
UNTIL DOOR IS OPENED
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
Screen with Service Instructions
Alarm
Description
Screen with Operator Instructions