12-4
129373-496
Operator Manual
Troubleshooting
Screen with Operator Instructions
Occurs if chamber
does not air break the
vacuum to 2 inHg
(0.07 Vbar) within the
allotted time.
Too Long In Evacuation
(Continued)
NOTE: This alarm has
two service help
screens.
12.2.4 Too Long In Air
Break
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
225
TOO LONG IN AIR BREAK
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
•
EXTEND AIR BREAK TIME
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
Alarm
Description
Screens with Service Instructions
EXIT
STATUS
PAPER
MORE
FEED
HELP
STATUS..
SERVICE INFORMATION:
203
TOO LONG IN EVACUATION
→
→
→
→
→
CHAMBER DID NOT REACH REQUIRED
VACUUM LEVEL WITHIN ALLOTTED TIME
CAUSES AND CORRECTION:
1.
WATER PRESSURE LESS THAN 30 psig
→
→
→
→
→
CHECK WATER SUPPLY PIPING
2.
CHAMBER DRAIN STRAINER PLUGGED
→
→
→
→
→
CLEAN
3.
CHECK VALVE MALFUNCTION
→
→
→
→
→
REPAIR
4.
SOLENOID VALVE MALFUNCTION
→
→
→
→
→
REPAIR
5.
DOOR SEAL NOT ACTIVATED
→
→
→
→
→
CHECK SEAL
→
→
→
→
→
CHECK SEAL STEAM AND EXHAUST
→
→
→
→
→
CHAMBER DID NOT REACH REQUIRED
VACUUM LEVEL WITHIN ALLOTTED TIME
CAUSES AND CORRECTION:
6.
LEAK IN PLUMBING
→
→
→
→
→
REPAIR
→
→
→
→
→
RUN A LEAK TEST
7.
CONTROL OUT OF CALIBRATION
→
→
→
→
→
RECALIBRATE CONTROL (CONTACT
QUALIFIED SERVICE PERSON)
STATUS..
SERVICE INFORMATION:
204
TOO LONG IN EVACUATION
PAPER
SERVICE
EXIT
FEED
MODE