10-23
Out of Cycle Options
Operating Instructions
129373-496
Pressing
←
or
→
on numeric keypad
moves the entry location to the left or
right, respectively.
Pressing CANCEL returns display to
screen #48 without changing the current
access code.
7.Enter the
new access code
using the numeric keypad. New access
code appears on display as it is being entered. Once the new code is
correctly entered, press
ENTER
.
84
CHANGE VALUES ACCESS CODE
1
2
3
4
5
6
7
8
9
←
←
←
←
←
0
→
→
→
→
→
CANCEL
ENTER THE NEW
ACCESS CODE.
ENTER
8.Screen #111 allows supervisor to designate which values can not be
changed by the operator. To lock out a specific value, press the
corresponding
values button
.
For example: Press
SETUP
to prevent the operator from changing the
setup values. Once pressed, the values button reads LOCKOUT.
NOTE: Pressing CYCLES advances display to screen #112. Screen
#112 allows supervisor to designate specific cycles which cannot be
changed by the operator.
Once all desired values are locked out, press
RETURN
.
111
LOCKOUT VALUES
TOO LONG
IN PHASE
TIME/
DATE
CYCLES
SETUP
RETURN